Development of a force balance telemetry device for total knee arthroplasty
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective A new type of tibialfemoral forcebalancing telemetry device was developed for measuring force balance on the tibialfemoral contact surface in knee joint. MethodThe forcebalance telemetry device was designed imitating the configuration of tibial part of total knee prosthesis, in which force sensors and telemetry circuit were embedded. During total knee arthroplasty(TKA), this device was implanted as a tibial prosthesis to measure tibialfemoral force while the knee joint was being placed at different angles. The tension balance was adjusted by releasing soft tissue around knee joint. ResultsThe results indicat that this device has advantage of high accuracy for force measurement (r>0.98,RMS= 65 g), which is sufficient to meet clinical demand. ConclusionsThe measurement of force balance on the tibialfemoral contact surface with telemetry device can provide new insight into the estimation of optimal knee prosthesis position and selection of appropriate polyethylene insert.

    Reference
    Related
    Cited by
Get Citation

DANG Xiao, ZHU Zhong-lin, DING Hui, ZHOU Yi-xin, WANG Guang-zhi. Development of a force balance telemetry device for total knee arthroplasty[J]. Journal of medical biomechanics,2010,25(2):112-117

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:January 19,2010
  • Revised:March 03,2010
  • Adopted:
  • Online:
  • Published: