BIOMECHANICAL COMPARISON STUDIES OF DIFFERENT INTERFIXATION FOR VERTICALLY UNSTABLE PELVIC FRACTURE
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    According to the biomechanical principle, 18 specimens of human pelvie were selected from freshcadavers and made into samples for biomechanical laboratory test. Using cadaveric pelvicinstrumented with strain gauges displacement instrument compare five different interfixationresults. Blomechanical studies demonstrated utilizing π- shaped rod and T - shaped plates,sufaficient stability of pelvic ring could be Provided. Its the Strength, rigidity and stability appear tooffer distinct biomechanical advantage over the three interfixation method (P<0.01). BiomechamIcal Parameters were provided the theoretical basis in clinical application of different interfixationfor Vertically unstable pelvic fracture.

    Reference
    Related
    Cited by
Get Citation

. BIOMECHANICAL COMPARISON STUDIES OF DIFFERENT INTERFIXATION FOR VERTICALLY UNSTABLE PELVIC FRACTURE[J]. Journal of medical biomechanics,1999,14(2):97-101

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online:
  • Published: