Stability of Fixation for AO-C2 Pilon Fracture with Horseshoe Ring External Fixator under Axial Loading
DOI:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective The traditional circular external fixation structure was redesigned to make it more suitable for intraarticular Pilon fracture fixation. The stability of the horseshoe annular external fixator was verified by analyzing fracture displacement of the Pilon bone after fixation with the external fixator under loading.MethodsAO-C2 Pilon fracture model was prepared by pendulum saw according to predetermined osteotomy line, and then fixed by horseshoe ring external fixator. The fixed ankle model was subjected to axial loading on mechanical test machine. The axial force loading mode was from 0 N to 150 N, 300 N and 450 N successively. The displacement changes of fracture blocks in X (left-right), Y (up-down) and Z (anterior-posterior) axis during axial loading were recorded by dynamic capture instrument.ResultsThe overall left-right, up-down and anterior-posterior displacement of the fibula ranged from 0.32 mm to 0.70 mm, -0.27 mm to -0.23 mm and 0.23mm to 0.32 mm, and the maximum difference was 0.09 mm. The overall left-right, up-down and anterior-posterior displacement of medial malleolus fracture in the tibia range from 0.02 mm to 0.14 mm, -0.80 mm to -0.19 mm and -0.78 mm to -0.13mm.The overall left-right, up-down and anterior-posterior displacement of lateral malleolus fracture in the tibia ranged from -0.07 mm to 0.05 mm, -0.36 mm to -0.03 mm, 0.27 mm to 0.47 mm.ConclusionsThe horseshoe ring external fixator can not only stabilize the fracture end, but also produce micromotion which is beneficial to fracture healing, which accords with biomechanical characteristics of fracture healing. The external fixator structure is simple, suitable for the treatment of Pilon fracture, and worthy of popularization and application.

    Reference
    Related
    Cited by
Get Citation

CHEN Yang, LIU Guangwei, CHENG Yongzhong, HUANG Xiaoyu, CHEN Yili, CHEN Hongying, WANG Xiongwei, ZHAO Jiyang, SHI Changlong. Stability of Fixation for AO-C2 Pilon Fracture with Horseshoe Ring External Fixator under Axial Loading[J]. Journal of medical biomechanics,2022,37(3):498-503

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 07,2021
  • Revised:August 22,2021
  • Adopted:
  • Online: June 24,2022
  • Published: