The Influence of Bending Parameters on Mechanical Properties of Soft Tissue Suture Passer Hook
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective To explore the influence of hook deflection angle and inclination angle on mechanical properties of the soft tissue suture passer hook. Methods Taking the end face far away from the tip of the needle (end face 1) as the study object, a mathematical model was established with the moment as dependent variable and the hook deflection angle and inclination angle as the independent variable. The moment was solved by the mathematical model with the deflection angle and inclination angle of 0°, 10°, 20° and 30°. Based on the finite element analysis method, 16 three-dimensional geometric hook models with deflection angle and inclination angle of 0°, 10°, 20° and 30° were established by SolidWorks. The stress analysis was carried out by ANSYS Workbench. Under the same puncture force, the maximum von Mises stress of each hook and the reaction moment of end face far away from the tip of the needle were calculated. Results The results from theoretical analysis and numerical simulation showed that the reaction moment of end face 1 increased with the increase of deflection angle, and increased with the decrease of inclination angle. The hook with deflection angle of 0° and inclination angle of 30° had the minimum reaction moment. The finite element analysis results showed that with the deflection angle of 0°, the maximum von Mises stress of the hook was the smallest and did not change with the inclination angle of the hook changing. Conclusions The established mathematical model can accurately explain the relationship between the moment at the end face of the hook and the deflection angle and inclination angle of the hook. This study provides the theoretical basis for designing hook geometry of the soft tissue suture passer, and improves the safety of the soft tissue suture passer in operation process.

    Reference
    Related
    Cited by
Get Citation

XIAO Shanshe, YUAN Shiting, GU Xuelian, LAI Weiguo, CHEN Zhi. The Influence of Bending Parameters on Mechanical Properties of Soft Tissue Suture Passer Hook[J]. Journal of medical biomechanics,2021,36(3):402-406

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 13,2020
  • Revised:July 07,2020
  • Adopted:
  • Online: July 05,2021
  • Published:
Article QR Code