Mechanical performance of targeted drug eluting stent during compression process
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective To investigate mechanical performances of two different drug eluting stents (DES) during compression process, so as to provide a scientific guideline for design of DES. Methods Using Solidworks, two kinds of stent models, namely the common DES without slots in outside surface (Stent I) and the targeted DES with slots in outside surface (Stent II) were established, respectively. The important mechanical properties of the two stents during compression process, such as rebound ratio of radial contraction, rebound ratio of axial contraction, the inhomogeneity of compression and the distribution of residual stress, were analyzed by ABAQUS. Results With the same outer radius, the difference values of rebound ratio of radial contraction, rebound ratio of axial contraction and inhomogeneity of compressing for the two stents were 0.07%-0.12%, 0.016%-0.033% and 0.013 %-0.048%, respectively; there was a bigger low-stress area on the support of Stent II. Conclusions Compared with the traditional DES, the targeted DES could maintain the same mechanical performances while reducing drug-loading capacity, which shows a good prospect in clinical application of the treatment for cardiovascular stenosis diseases.

    Reference
    Related
    Cited by
Get Citation

GUO Jing-zhen, SONG Cheng-li, CUI Hai-po. Mechanical performance of targeted drug eluting stent during compression process[J]. Journal of medical biomechanics,2014,29(6):524-529

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:January 13,2014
  • Revised:March 17,2014
  • Adopted:
  • Online: December 24,2014
  • Published:
Article QR Code