Objective To study the flow field of a new vascular bioreactor with small diameter at different operating conditions. Method To investigate the flow field of a new vascular bioreactor with small diameter under three operating conditions by numerical simulation: outerbox rotating, innerbox rotating and both innerbox and outerbox rotating. A comparison of parameters such as flow velocity and shear stress was made. Results The culture medium in the bioreactor rotated with the innerbox and outerbox, and the velocity was uniformly distributed; the bioreactor could therefore provide a low shear stress environment that is nondestructive for the cells in the medium; the shear stress in bioreactor had a quadratic function relation with the rotation rate. Conclusions This bioreactor could provide a good culture environment for vascular tissue regeneration, and this study provided theoretical basis for the experimental research of bioreactors.
Reference
Related
Cited by
Get Citation
ZHU Sujing, ZHENG Tinghui, ZOU Yuanwen, LI Jinchuan, HUANG Xuejin, FAN Yubo. Numerical analysis on flow field in a vascular bioreactor[J]. Journal of medical biomechanics,2010,25(5):352-356