    Objective we carried out a biomechanical comparison of a new point contact locking compression plate and the dynamical compression plate in in an osteoporotic comminuted radial fracture model. Methods Twelve pairs of aged osteoporotic cadaveric radii, 1cm mid-shaft transverse fracture were created to simulate comminuted fracture and fixed with PC-LCP、DCP, respectively. Then the anteroposterior four point bending、mediolateral four point bending and torsion tests was conducted to determine the bending、and torsion compress stiffness of the PC-LCP and DCP. Results Although both the bending and torsion stiffness of PC-LCP are higher than that of DCP, there are no significant differences. Conclusion The new PC-LCP has a tendency of the higher bending and torsion stiffness when used to fix the steoporotic comminuted radial fracture.

    Reference
    Related
    Cited by
Get Citation

XIONG Yan, ZHAO Yu-feng, WANG Zi-ming, DU Quan-yin, SUN Hong-zhen, WU Si-yu, CHEN Wei-jun, WANG Ai-ming, YU Ming-an, YANG Yu-xiang. Biomechanical research on a new point contact locking compression plate in an osteoporotic comminuted radial fracture model[J]. Journal of medical biomechanics,2009,24(3):200-203