Biomechanical research on a new designed point compress locking plate
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective Design a newtype bone plate PC-LCP and determine its mechanical properties and the biomechanical stability.Method Four point bending test was conducted to determine the mechanical properties of each of the 3 pairs of PC-LCP、LC-DCP、DCP;using 20 pairs of new fresh sheep tibia,mid-shaft transverse fracture were created and fixed with PC-LCP、LC-DCP、DCP,respectively;and then the four point bending、torsion and axial compress tests were conducted to determine the bending、torsion and axial compress stiffness.Result Although the bending stiffness of PC-LCP is less than that of DCP,LC-DCP and have significant difference,the biomechanical properties of fixed fracture bone by these three type bone plate showed no significant difference.Conclusion The new PC-LCP has the similar biomechanical properties as the other two type plates when used to fix the broken bone and has the adequate stability.

    Reference
    Related
    Cited by
Get Citation

. Biomechanical research on a new designed point compress locking plate[J]. Journal of medical biomechanics,2007,22(3):290-293