Biomechanic research on ZH transpedicular internal fixator used for thoracolumbar fracture treatment
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective To provide the biomechanical basis for ZH transpedicular internal fixator used in thoracolumbarfracture treatment. Methods Trauma models were made from 9 fresh cadavers by Panjabi method and divided into 3 groupswith ZH transpedicular internal fixator, Dick screw and Steffee plate, respectively. To test their biomechanical capability inload-responsive change、centrum intensity、displacement、torsion intensity、stability, rigidity and ultimate strength.Results The results of the test showed that ZH transpedicular internal fixator was much better than the other two oneswith: load-responsive change was much less than that of Dick screw and Steffee plate by 7.2% and 14% at Point A andby 17% and 28% at Point B; Centrum intensity: stress decreased by 7% and 14% at Point A, while at Point B by 17% and29%; The displacement decreased by 24% and 33% vertically, and 40% and 55% horizontally. Conclusions The ZHtranspedicular internal fixator is suitable to popularized in clinic application with its better biomechanical capability.

    Reference
    Related
    Cited by
Get Citation

. Biomechanic research on ZH transpedicular internal fixator used for thoracolumbar fracture treatment[J]. Journal of medical biomechanics,2006,21(3):235-240

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online:
  • Published:
Article QR Code