BIOMECHANICAL ANALYSIS OF SUTURE LINE STRESS IN THE END-TO-SIDE ANASTOMOSIS
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    The tensile force acting on the suture line connecting a synthetic graft to a host artery in end-to-side anastomosis is shown to increase as the ratio of the graft-artery area,and has minimum value at 0.3 of the ratio of the flow rates.The angle of the end-to-side anastomosis is shown to have little effect on the suture line stress,but can not be negligated.Considering suture material and other factors,sufficient sutures per cm should be established.

    Reference
    Related
    Cited by
Get Citation

. BIOMECHANICAL ANALYSIS OF SUTURE LINE STRESS IN THE END-TO-SIDE ANASTOMOSIS[J]. Journal of medical biomechanics,1995,10(4):205-208