新型射频组织焊接电极的设计、仿真与实验研究
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

科技部国家重点研发计划项目(2019YFC0120402),国家自然科学基金项目(51901137,51735003)


Design, Simulation and Experimental Study of a Novel Radiofrequency Tissue Welding Electrode
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 设计一种新型射频组织焊接电极,提高吻合口生物力学强度的同时减少组织热损伤。 方法 设计表面存在镂空结构的新型电极(梅花形电极),以环形电极作为对照组,在射频能量作用下完成肠道组织焊接,通过撕脱力和爆破压测试研究焊接吻合口的生物力学特性,采用有限元电-热-力多场耦合仿真分析和热电偶探针研究焊 接过程中的组织热损伤,并对微观组织结构进行检查。 结果 当焊接功率 120 W、焊接时间 8 s、压合压强 20 kPa时,肠道吻合口呈现最优的生物力学特性。 与环形电极对照组相比,梅花形电极组吻合口生物力学强度更高,撕脱力和爆破压分别从(8. 62±1. 22) N、(81. 7±3. 36) mmHg 增加到 (9. 54±1. 24) N、(89. 4±4. 15) mmHg,且组织热损伤显著减少,组织微观结构连接更为紧密。 结论 该新型电极在提高吻合口生物力学强度的同时可减少组织热损伤,进而实现更好的吻合效果。 研究结果可为实现人体管腔组织的无缝连接提供参考。

    Abstract:

    Objective Aiming at improving biomechanical strength of the anastomotic stoma as well as reducing tissue thermal damage, a novel radiofrequency (RF) tissue welding electrode was developed. Methods A novel electrode with a hollow structure on the surface ( the plum electrode) was designed and the ring electrode was used as control group to conduct the welding of intestinal tissues based on RF energy. Biomechanical properties of anastomotic stoma were studied by shear test and burst pressure test. The tissue thermal damage during welding was investigated by finite element electro-thermal-mechanical multi-field coupling simulation analysis and thermocouple probe, and the tissue microstructures were also studied. Results Under 120 W RF energy, 8 s welding duration and 20 kPa compression pressure, the anastomotic stoma had the optimal biomechanical properties. Compared with the ring electrode group, biomechanical strength of the anastomotic stoma in plum electrode group was higher, with the shear strength and burst pressure increasing from (9. 7±1. 47) N, (84. 0±5. 99) mmHg to (11. 1±1. 71) N, (89. 4±6. 60) mmHg, respectively. There was a significant reduction in tissue thermal damage, and intact and fully fused stomas could be formed in anastomotic area. Conclusions The proposed novel electrode could improve biomechanical strength of the anastomosis as well as reduce tissue hermal damage, thus achieve better fusion. The research result provide references for realizing the seamless connection of human lumen tissues

    参考文献
    相似文献
    引证文献
引用本文

邢绪坡,胡钟欣,韩正一,宋成利,毛琳.新型射频组织焊接电极的设计、仿真与实验研究[J].医用生物力学,2023,38(3):601-607

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2022-07-02
  • 最后修改日期:2022-07-27
  • 录用日期:
  • 在线发布日期: 2023-06-27
  • 出版日期:
文章二维码
关闭