心室内隧道的补片形状设计及有限元分析
DOI:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

国家自然科学基金项目(11772015),北京市医院管理局临床技术创新项目2018年度“扬帆”计划(XMLX201845)


Structure Design and Finite Element Analysis on Patch in Intraventricular Tunnel
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 探究心室内隧道(intraventricular tunnel, IVT)构建术的补片形状对右心室双出口矫治术力学性能的影响。方法 基于理想化IVT模型,设计二维IVT补片。根据转弯部分菱形补片长短轴不同,建立长短轴比分别为1∶0.625、1∶0.3、1∶0.2、1∶0.15、1∶0.125、1∶0.1共计6组补片模型,利用有限元分析方法对补片缝合、保持、撑起为三维IVT模型的过程进行数值模拟。结果 6组补片模型的缝合缘最大应力主要集中在转弯部分菱形锐角边缝合处;随着补片长短轴比增大,IVT缝合缘最大应力先减少后增大,体积呈现增大的趋势,隧道两端压差呈现先减小后增大的趋势。菱形长短轴比为1∶0.15补片的表面应力分布均匀,且缝合缘上最大应力最小,同时对右心室体积侵占较小,隧道两端的压差较小。结论 IVT补片形状对IVT缝合缘应力、右心室体积及隧道压差有影响,而且不是单调变化。菱形长短轴比为1∶0.15的补片缝合效果在所建模型中相对较好。

    Abstract:

    ObjectiveT o explore the influence of patch shape for intraventricular tunnel (IVT) construction on biomechanical performance of the double outlet right ventricle after correction. Methods Based on the idealized IVT model, a two-dimensional IVT patch was designed. Six groups of patch models with the rhombic long-to-short axis ratio of 1∶0.625, 1∶0.3, 1∶0.2, 1∶0.15, 1∶0.125, 1∶0.1 were established according to the difference between the long and short axis of the rhombus patch in the turning part, and finite element analysis method was used to numerically simulate the process of stitching, holding and propping up the patch into a three-dimensional (3D) IVT model. Results The maximum stresses on suture line of 6 patch models were mainly concentrated at acute-angle corners of the rhombus. As rhombic long-to-short axis ratio of the patch increased, the maximum stress of the IVT suture line first decreased and then increased, and the volume showed an increasing trend. The pressure difference between two ends of the tunnel first decreased and then increased. The patch with the long-to-short axis ratio of 1∶0.15 had a uniform surface stress distribution, and the maximum stress on the suture line was the smallest. Meanwhile the right ventricular volume was less encroached on, and the pressure difference at both ends of the tunnel was small. Conclusions The IVT shape can influence stresses of suture line, the right ventricle volume and the pressure difference of IVT with non-monotonic variations. The suture effect of the patch with the long-to-short axis ratio of 1∶0.15 is relatively better among the constructed models.

    参考文献
    相似文献
    引证文献
引用本文

李晓,王俊杰,柳思聪,刘宁,杨印龙,杨尧,范祥明,乔爱科.心室内隧道的补片形状设计及有限元分析[J].医用生物力学,2022,37(1):98-104

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-12-20
  • 最后修改日期:2021-03-18
  • 录用日期:
  • 在线发布日期: 2022-02-25
  • 出版日期:
文章二维码
关闭