顺应性测试条件对纺织人工血管径向顺应性的影响
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

国家教育部博士点基金项目(20100075110001),高等学校学科创新引智计划资助项目(B07024)


Effects of testing conditions on radial compliance of textile vascular prostheses
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 分析纺织人工血管试样的预加张力和测试位置,探讨影响人工血管径向顺应性的因素。方法 基于人工血管动态顺应性测试仪,通过变化试样轴向预加张力以及调整试样直径测试观察点位置,对3种内直径8 mm的纺织人工血管试样,测试不同脉动压力值及其所对应的径向直径值。根据ISO 7198-1998顺应性计算方法,获得不同实验条件下的径向顺应性值。结果 随着预加张力的增大,3种纺织人工血管径向顺应性也随之增大,当预加张力达到500~550 mN时,径向顺应性值达到最大(1.03~1.72)%/100 mmHg(1 mmHg=0.133 kPa),之后随着预加张力的增大而降低。此外,PET波纹化经编人工血管的径向顺应性值显示波谷径向顺应性值略大于波峰,而波峰的径向顺应性值稳定性好于波谷。结论 预加张力和测试位置对纺织人工血管径向顺应性有较大的影响,在测试时应给试样的预加张力为500~550 mN,对于波纹化纺织人工血管应选择波峰处为测试位置,结果更加稳定,数据更加可靠。

    Abstract:

    Objective To explore influences of pre-tension and testing site on the radial compliance of vascular prostheses. Methods Three different textile vascular prosthesis samples (PET woven, PET/PPT woven and PET corrugate warp knitted) with the diameter of 8 mm were selected. Under various pre-tensions and testing sites, the pulsatile pressure and its corresponding radial diameter of these samples were measured by dynamic compliance testing instrument. The radial compliance was calculated according to the compliance formula in ISO 7198-1998. Results The radial compliance of all the three vascular prostheses increased along with the pre-tension increasing. When the pre-tension was 500~550 mN, all the radial compliance reached the maximum (1.03~1.72)%/100 mmHg (1 mmHg=0.133 kPa), and then began to decrease with pre-tension increasing. Furthermore, the radial compliance of corrugate vascular prostheses was slightly bigger at valleys than that at peaks, while the radial compliance at peaks was more stable than that at valleys. Conclusions The pre-tension and test site had vital effects on testing the radial compliance of vascular prostheses. Pre-tension of 500~550 mN should be right to apply to all the three kinds of textile vascular prosthesis at test. As for corrugate vascular prostheses, the radial compliance at peaks should be chosen as the test site to get more stable result and reliable data.

    参考文献
    相似文献
    引证文献
引用本文

沈高天,张蔚,林婧,王璐.顺应性测试条件对纺织人工血管径向顺应性的影响[J].医用生物力学,2014,29(2):113-118

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2013-03-20
  • 最后修改日期:2013-03-30
  • 录用日期:
  • 在线发布日期: 2014-04-18
  • 出版日期:
文章二维码
关闭