新型股骨内髁逆行髓内针内固定治疗股骨远端A型骨折的有限元分析
DOI:
CSTR:
作者:
作者单位:

1.福建医科大学附属闽东医院;2.福建医科大学

作者简介:

通讯作者:

中图分类号:

基金项目:


Finite element analysis of a new retrograde intramedullary nail on medial femur condyle for internal fixation on distal type A femur fracture
Author:
Affiliation:

1.Mindong Hospital affiliated to Fujian Medical University;2.Fujian Medical University

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 采用有限元分析比较新型股骨内髁逆行髓内针、股骨普通逆行髓内针、股骨内髁锁定钢板治疗股骨远端A型骨折的生物力学特点,研究股骨内髁逆行髓内针的优势 。方法 设计新型股骨内髁逆行髓内针。CT扫描1名男性志愿者下肢骨,通过软件建立股骨三维模型,对模型进行分割,建立股骨远端A2、A3型骨折模型。建立内固定物三维模型。A2骨折采用股骨内髁逆行小型髓内针(小型组)和标准型髓内针(标准组),普通逆行髓内针(普通组),股骨内髁锁定钢板(钢板组)内固定。A3骨折分别采用上述一种内固定物联合外髁钢板内固定 。对A2、A3骨折模型施加600 N轴向载荷及4N·m扭转载荷 ,观察各组模型中股骨位移、应力,内固定物位移、应力,内固定物与骨骼间微动量,骨折两断端相对位移。结果 A2、A3骨折轴向载荷中标准(联合)组股骨和内固定物移位峰值、应力峰值最小。A2、A3骨折扭转载荷中标准(联合)组股骨和内固定物位移峰值最小。骨折两断端相对位移中标准联合组最小。结论 对比股骨内髁锁定钢板和普通逆行髓内针,新型股骨内髁逆行髓内针具有减少应力集中,降低内固定物失效风险的力学优势,可单用于股骨远端A2型骨折,或联合股骨外髁钢板固定股骨远端A3型骨折。

    Abstract:

    Objective To compare the biomechanical characteristics of the new retrograde intramedullary nail, common femoral retrograde intramedullary nail, and medial femoral condyle locking plate in the treatment of type A distal femoral fractures by finite element analysis, and to study the advantages of retrograde intramedullary nail on medial femoral condyle.Methods Design a new new retrograde intramedullary nail on medial femur condyle. A CT scan was performed on the lower limb bone of a male volunteer, and a three-dimensional model of the femur was established using soft. The model was then segmented to create a three-dimensional model of a femoral distal A2、A3 type fracture. The three dimensional models of internal fixator were established. The small and standard retrograde intramedullary nailing on medial femoral condyle, common femoral retrograde intramedullary nailing, locking plate on medial femoral condyle were used for internal fixation on A2 type fracture. A3 type fracture was fixed with one of the above internal fixators combined with lateral condylar plate respectively.The axial load of 600 N and the torsional load of 4N·m were applied to the models. The displacement and stress of femurs , the displacement and stress of internal fixators, the micro-momentum between the internal fixator and the bone, the relative displacement of the two fracture broken ends were observed in each group. Results Among the A2 and A3 type fractures under axial loads, the standard (combined) group had the smallest peak displacement and peak stress of the femur and internal fixation. Among the A2 and A3 type fractures under torsional loads, the standard group (combined) had the smallest peak displacement of the femur and internal fixation. The relative displacement of the two broken ends in the standard combined group was the smallest. Conclusion Compared with femoral medial condylar locking plate and common retrograde intramedullary needle, the new retrograde intramedullary needle on medial femur condyle has mechanical advantages of reducing stress concentration and reducing the risk of internal fixation failure,can be used alone for distal femoral A2 type fracture, or in combination with the lateral condylar plate to fix distal femoral A3 type fracture.

    参考文献
    相似文献
    引证文献
引用本文
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2025-01-09
  • 最后修改日期:2025-02-09
  • 录用日期:2025-02-20
  • 在线发布日期:
  • 出版日期:
文章二维码
关闭