不同类型主动脉瓣偏心放置下瓣后流场特性的粒子图像测速实验研究
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

中国医学科学院医学与健康科技创新工程(2021-I2M-1-006),国家自然科学基金项目(12402382)


Flow Field Characteristics of Aortic Valve with Eccentric Lower Valve Placement:A PIV Experimental Study
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 探究不同类型人工主动脉瓣膜的偏心放置对瓣后流场的影响。方法 建立体外脉动循环模拟系统,利用粒子图像测速(particle image velocimetry,PIV)实验分别研究生物瓣和机械瓣在居中放置(0 mm)及偏心放置(1 mm)两种情况下瓣后流场变化,通过速度、涡度及黏性剪切应力血流动力学参数对流场状态进行评估。结果 通过分析瓣膜在收缩早期、加速期、峰值期及减速期4个特征时刻的流场变化,发现生物瓣和机械瓣的流场分布存在显著差异,生物瓣为中心射流左右对称,流速较大;机械瓣呈现三孔射流,中心流速较低。在偏心放置的情况下,两种瓣膜窦部区域血流缓慢,平均流速降低,不利于涡流的形成和维持。心动周期峰值期生物瓣和机械瓣窦部最大黏性剪切应力分别为0.45、0.67 N/ m2,接近内皮细胞损伤阈值。结论 机械瓣和生物瓣在偏心放置情况下会导致窦部血流速度降低,黏性剪切应力减小,从而为血栓形成创造了有利条件。临床手术中应注意瓣膜置换的位置,避免出现偏心放置的情况。

    Abstract:

    Objective To investigate the impact of eccentric placement for various types of artificial aortic valves on downstream flow dynamics. Methods A physiological pulsatile circulation simulation system was emoloyed and particle image velocimetry (PIV) was utilized to analyze the downstream flow field variations for bioprosthetic and mechanical valves under two placement conditions: centralized placement (0 mm) and eccentric placement (1 mm). Hemodynamic parameters such as velocity, vorticity, and viscous shear stress were assessed to evaluate the flow field characteristics. Results By analyzing the flow field variations at four characteristic time points, namely, early systole, acceleration phase, peak systole, and deceleration phase, a significant difference in flow field distribution between bioprosthetic and mechanical valves was observed. The bioprosthetic valve exhibited a centrally symmetric jet with higher flow velocity, whereas the mechanical valve displayed a three-jet structure with a lower central flow velocity. Under eccentric placement, the blood flow in the aortic sinus region was sluggish, with a reduction in average velocity, hindering the formation and maintenance of vortices. During the peak systolic phase, the maximum viscous shear stresses in the sinus region for the bioprosthetic and mechanical valves were 0.45 and 0.67 N/m2, respectively, approaching the threshold for endothelial cell damage. Conclusions Eccentric placement of both mechanical and bioprosthetic valves resulted in reduced sinus blood flow velocity and diminished viscous shear stress, creating favorable conditions for thrombus formation. In clinical practice, careful attention should be given to the placement of valve replacements to prevent eccentric placement.

    参考文献
    相似文献
    引证文献
引用本文

韩恩慧,侯倩文,肖洋,孟雅娜,魏海洋,江誉,胡建军,周建业.不同类型主动脉瓣偏心放置下瓣后流场特性的粒子图像测速实验研究[J].医用生物力学,2025,40(1):25-33

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2024-10-17
  • 最后修改日期:2024-12-03
  • 录用日期:
  • 在线发布日期: 2025-02-26
  • 出版日期: