    摘要:

    目的 分析不同结构的取栓支架的机械性能,为取栓支架的设计以及临床选择提供理论支持。方法 通过有限元分析以及体外模拟实验对3款不同结构的取栓支架(B3、K4、X)进行评估,分析指标是径向支撑力、回撤力以及模拟取栓过程中的血栓状态。结果 实验得到B3、K4、X取栓支架的径向支撑力分别为0.48、0.43、0.51 N,且压握尺寸越大,取栓支架的径向支撑力越大;当压握尺寸大于3 mm时,支架的径向支撑力显著增大。模拟取栓实验结果表明,B3、K4、X支架的回撤力峰值分别为0.410、0.451、0.501 N,实验结果与有限元分析的结果具有趋势一致性。结论 实验结果与有限元结果均表明,X支架具有更好的力学性能。此方法可作为评估取栓支架性能的一种分析方法,为取栓支架的性能提高和研发提供参考。

    Abstract:

    Objective To analyze mechanical properties of stent retrievers with different structures, and provide theoretical support for the design and clinical selection of stent retrievers. Methods Three kinds of stent retrievers with different structures (B3, K4, X) were evaluated by finite element analysis and in vitro simulation experiment. The analytic parameters were radial support force, withdrawal force and thrombosis state during the thrombotomy test. Results The radial support forces of B3, K4, X stent retrievers obtained from the experiment were 0.48 N, 0.43 N, 0.51 N, respectively. The larger the crimping distance, the greater the radial support force of stent retrievers. The radial support force of the stent increased significantly when the crimping distance was larger than 3 mm. The simulated thrombus removal experiment results showed that the peak withdrawal forces of B3, K4, X stent retrievers were 0.410 N, 0.451N, 0.501 N, respectively. The experimental results were consistent with the finite element analysis results. Conclusion sBoth the experimental results and the finite element results showed that the X stent has better mechanical properties. This method can be used as an analytic method to evaluate performance of the stent retrievers, and provide references for performance improvement and development of the stent retrievers.

    参考文献
    相似文献
    引证文献
引用本文

邱晓键,张晓颖,谷雪莲,肖有申,赵宇轩.优化取栓支架的力学性能[J].医用生物力学,2022,37(3):419-424

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2021-06-11
  • 最后修改日期:2021-07-20
  • 录用日期:
  • 在线发布日期: 2022-06-24
  • 出版日期:
文章二维码
关闭