    摘要:

    目的 研究当前临床常用的不同材料匹配的典型人工髋关节接触力学性能。方法 建立球基共型接触人工髋关节的有限元模型;通过施加恒定竖直方向载荷,等效模拟关节正常行走情况下的接触力学,分析材料匹配球面的接触性能及评估相应的磨损预测。结果 获得具有诸如金属对金属(MOM)、金属对陶瓷(MOC)、陶瓷对陶瓷(COC)、陶瓷对聚合物(COP)、金属对聚合物(MOP)5种材料组配条件下的人工髋关节接触性能数值分析;结果显示其接触应力由大到小排列分别为COC、MOC、MOM、COP和MOP;而接触面积由大到小排列分别为MOP、COP、MOM、MOC和COC。结论 高弹性模量和低泊松比的材料匹配的人工髋关节变形量较小,但会导致局部应力较大;将低弹性模量和高泊松比的柔性材料聚乙烯作为髋臼,则不会出现较大的接触应力集中,但变形量较大,且出现边缘受力情况。另一方面,MOM和MOC材料匹配的人工髋关节具有较低的接触磨损性能,对临床上人工髋关节选材提供参考价值。

    Abstract:

    Objective To study the contact performance of hip joint replacements with different material combinations typically used in clinic. Methods The finite element model of spherical conformal contact of hip joint replacements was developed to analyze the different contact mechanics performance of hip joint replacements with different material combinations by simulation test under a vertical constant load within one normal walking gait. The corresponding wear prediction was also evaluated. Results The contact mechanics performances of hip joint replacements with the material combination of metal-on-metal (MOM), metal-on-ceramic (MOC), ceramic-on-ceramic (COC), metal-on-polyethylene (MOP) and ceramic-on-polythene (COP) were analyzed. The results obtained from the simulations showed that contact stress was decreased in the order of COC>MOC>MOM>COP>MOP, while contact area was decreased in the order of MOP>COP>MOM>MOC>COC. Conclusions Hip joint replacement with the material combination of high elastic modulus and low Poisson’s ratio has smaller deformation, but could induce larger local stress, while flexible material polythene of low elastic modulus and high Poisson’s ratio might have smaller concentrated contact stress, but cause larger deformation and edge contact phenomenon on the acetabulum cup of hip joint replacements. In addition, hip joint replacements with MOC and MOM material combination may have lower contact bearing performance, which provides references for material choices of typical artificial hip joints in clinic.

    参考文献
    相似文献
    引证文献
引用本文

毛锋,郭海洋,程倩,奚卉,张茜,王凤才.材料匹配与人工髋关节接触性能分析[J].医用生物力学,2014,29(1):38-45

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2012-12-05
  • 最后修改日期:2013-03-02
  • 录用日期:
  • 在线发布日期: 2014-02-24
  • 出版日期:
文章二维码
关闭