    摘要:

    目的 探索骨组织在微动过程中裂纹产生和扩展的原因及过程,弄清假体-骨组织界面微动磨损的机理。方法 用干燥股骨和钛合金构造成微动实验系统进行实验研究。结果 微动区域附近的骨组织均变疏松,纵剖面和横剖面上都存在肉眼可见的明显裂纹。结论 骨组织在微动过程中将发生疲劳破坏,从而造成假体松动。

    Abstract:

    Objective To investigate the cause and process of crack generation and propagation in fretting wear of bone and to study the fretting wear mechanism between prostheses and bones. Methods Conducting a fretting wear experiment system with a dried thigh bone and titanium alloy. Results Bone structures in fretting areas become loose and obvious cracks could be found in both longitudinal and cross -sectional planes of the bone. Conclusion Bone structures in fretting wear will induce fatigue failure and this will further cause loosening of prostheses.

    参考文献
    相似文献
    引证文献
引用本文

王野平,李望云,张菊华.假体-骨组织界面微动磨损机理的实验研究[J].医用生物力学,2003,18(1):23-27